Hipot Electrical Safety Ground Bond Analyzer





 
  Home
  About Us
  Need a Quote
  We Buy Equipment
  Used Equipment
  Reconditioned
  Rentals
  Repairs
  Leasing
  Payment Options
  Government
  Contact Us
  Home

19052 AC/DC Hipot / Insulation resistance (IR)
CE

19053 AC/DC Hipot / Insulation resistance (IR) / 8HV Channel Scan
UL, TUV, CE

Hipot Electrical Safety Ground Bond Analyzer

Specifications


Model 19071 19073 19052 19053
Mode AC AC/DC/IR AC/DC/IR AC/DC/IR/Scan
Scanner Unit - - - 8 ports,±phase
Withstanding Voltage Test
Output Voltage AC : 0.05 ~ 5kV AC : 0.05 ~ 5kV , DC:0.05 ~ 6kV
Load Regulation 1% of setting + 5V
Voltage Regulation 2V
Voltage Accuracy 1% of setting + 5 count
Cutoff Current AC : 0.1~20mA, DC : 0.01 ~ 5mA AC : 0.1~30mA, DC : 0.01 ~ 10mA
Current Resolution AC : 1µA, DC : 0.1µA
Current Accuracy 1% of setting + 5 count
Output Frequency 50Hz / 60Hz
Test Time 0.3 ~ 999 sec , continue
Ramp Time 0.1 ~ 99.9 sec, Off
Fall Time 0.1 ~ 99.9 sec, Off
Waveform Sine wave
Insulation Resistance Test
Output Voltage - DC : 0.05 ~ 1kV
Voltage Resolution - 2V
Voltage Accuracy - 5% of setting + 5 count 1.5% of setting + 5 count 1.5% of setting + 5 count
IR Range - 1MΩ ~ 50GΩ 1MΩ ~ 50GΩ 1MΩ ~ 10GΩ
Resistance Resolution - 0.1MΩ 0.1MΩ 0.1MΩ
Resistance Accuracy ≥500V 1MΩ ~1GΩ - 4% of setting + 5 count 5% of setting + 5 count 5% of setting + 5 count
1GΩ ~10GΩ - 7% of setting + 5 count 10% of setting + 5 count 10% of setting + 5 count
10GΩ ~50GΩ - 12% of setting + 5 count 15% of setting + 5 count 15% of setting + 5 count
≤500V 1MΩ ~1GΩ - 7% of setting + 5 count 10% of setting + 5 count 10% of setting + 5 count
≤100V 1MΩ ~1GΩ - 10% of setting + 5 count 10% of setting + 5 count 10% of setting + 5 count
Flashover (ARC) Detection
Setting Mode Programmable setting
Detection Current AC : 1mA ~ 15mA, DC : 1mA ~ 5mA AC : 1mA ~ 15mA, DC : 1mA ~ 10mA
Secure Protection Function
Fast Output Cut-Off 0.4ms after NG
Ground Fault Interrupt 0.5mA ±0.25mA AC, ON/OFF
Panel Operation Lock Present password
Continuity Check 1Ω ± 0.2Ω, ON/OFF
GO/NG Judgment Window
Indication,Alarm GO: Short sound,Green LED; NG: Long sound, Red LED
Data Hold Least tests data memories
Memory Storage 50 instrument setups with up to 20 test steps
Interface
Remote control Input : Start, Stop, Interlock (at 11 pin terminal block only) ; Output : Under test, Pass, Fail
Communication Interface RS485 (Option) RS232 (Standard), GPIB (Option)
General
Operation Environment Temperature : 0°C~40°C, Humidity : 15% to 95% R.H@≤40°C
Power Consumption 300W 500W
Power Requirements 100V/120V/220V/240V (AC ±10%), 50/60Hz
Weight Approx. 12KG Approx. 15KG
Certifications UL, TUV, CE UL, TUV, CE CE UL, TUV, CE

 

Hipot Electrical Safety Ground Bond Analyzer

19035 AC/DC Hipot, IR, DCR
CE Mark

The 19035 is designed to directly connect to a DWX Series Impulse Winding Tester and use the 8-Channel scanning capabilities to reach multiple points in one test instead of switching test points manually.

Specifications


Model 19035 19035-L 19035-M 19035-ML 19035-S
Mode AC/DC/IR/
DCR -8CH/ IWT
AC/DC/IR/ DCR -8CH/ IWT AC/DC/IR/
DCR -16CH/ IWT
AC/DC/IR/
DCR -16CH/ IWT
AC/DCR -8CH
Impulse Winding Test
(Layer Short, IWT)
External option Internal standard External option Internal standard -
Channel Programming H/L/X in 8CHs H/L/X in 8CHs H/X in CH 1,2,3,5,6,7
L/X in CH 4,8
H/X in CH 1,2,3,5,6,7
L/X in CH 4,8
H/L/X in 8CHs
Withstanding Voltage Test
Output Voltage

AC: 0.05~5KV / DC: 0.05~6KV

Load Regulation ±(1% setting + 0.1%full Scale)
Voltage Regulation 2V
Voltage Accuracy 1% setting + 0.1% full Scale
Cutoff Current

AC: 30mA / DC: 10mA

Current Resolution AC: 1µA / DC: 0.1µA
Current Accuracy 1% of reading + 0.5% of range (1% of reading + 5% of total current)
Current Frequency 50Hz / 60Hz
Test Time 0.1 ~ 999 sec , continue
Test / Ramp / Fall / Dwell Time 0.3 ~ 999 sec., continue / 0.1 ~ 999 sec., off / 0.1 ~ 999 sec., off / 0.1 ~ 999 sec., off
Wave Form Sine wave
Insulation Resistance Test
Output Voltage DC : 0.05 ~ 5kV
Voltage Resolution 2V
Voltage Accuracy 1% of setting + 0.1% of full range
IR Range 1MΩ ~ 50GΩ
Resistance Resolution 0.1MΩ
Resistance Accuracy ≧1000V 0.1MΩ ~ 1GΩ : ± (3% of reading + 0.1% of full range)
1GΩ ~ 10GΩ : ± (7% of reading + 2% of full range)
10GΩ ~ 50GΩ : ± (10% of reading + 1% of full range)
500V~1000V 0.1MΩ ~ 1GΩ : ± (3% of reading + 0.1% of full range)
1GΩ ~ 10GΩ : ± (7% of reading + 2% of full range)
10GΩ ~ 50GΩ : ± (10% of reading + 1% of full range)
≦500V 0.1MΩ ~ 1GΩ : ± 3% of reading + (0.2*500/Vs)% of full scale
DC Resistance Measurement
Test Signal <DC 10V. < DC 100mA
Measurement mode 2 terminals (2W) / 4 terminals(4W) measurement selectable
Measurement Accuracy
(2W/ 4W)
1Ω (4W only) ±(0.5% of reading + 0.5% of range)
10kΩ ± (2% of reading + 0.5% of range) / ± (0.5% of reading + 0.05% of range)
100kΩ
1kΩ
10kΩ
100kΩ
Flashover Detection
Setting Mode Programmable setting
Detection Current AC : 1mA ~ 15mA, DC : 1mA ~ 10mA
Secure Protection Function
Fast Output Cut-off 0.4ms after NG happen
Ground Fault Interrupt
0.5mA ±0.25mA AC, ON/OFF
Panel Operation Lock Present password
Interlock YES
GO/NG Judgment Window
Indication, Alarm
GO : Short sound, Green LED; NG : Long sound, Red LED
Data Hold Least tests data memories
Memory Storage 50 instrument setups with up to 20 test steps
Interface
RS232*1 (Standard), RS232*1 or GPIB & Handler & Temperature interface (Optional).
General
Operation Environment Temperature : 0°C~45°C, Humidity : 15% to 95% R.H@≦40°C
Power Consumption 500VA
Power Requirements 90~132Vac or 198~264Vac, 47~66Hz
Weight Approx. 20KG

 

The 19032 series Safety Analyzer able to perform several dielectric tests in one unit including AC/DC Hipot, Insulation Resistance,Ground Bond, Open/Short Check and optional Leakage Current for compliance testing of IEC, UL, TUV, CSA, EN and other standards including IEC60601-1, UL2601-1, IEC950, UL544, UL60950. The time-saving TwinPort™ feature performs Hipot and Ground Bond simultaneously cutting time for these tests in half.  The 19032's open / short check function (OSC) checks for an open connection or a DUT short during testing.  Open connections may cause a bad DUT to test OK.  Screening for shorts prevents damage to the equipment saving potential costs that may occur.  These patented features are only available in safety products manufactured by Chroma.

 

Hipot Electrical Safety Ground Bond Analyzer
Model 19032 19032-P (Coming soon)
Difference at a glance --- GFI Human Protection / Floating Curcuit Funtion / Mass Power Hipot Output Test 500VA
Softpanels (software) Yes Yes
Mode AC / DC / IR / GB / LC
Withstanding Voltage Test
Output Voltage DC:0.05 ~ 6kV , AC : 0.05 ~ 5kV
Load Regulation ±(1% of reading +0.1% of range) ±(1% of reading +0.1% of range)
Voltage Regulation 2V
Voltage Accuracy ±(1% of reading +0.1% of range) ±(1% of reading +0.1% of range)
Cutoff Current DC : 20mA , AC : 40mA DC : 20mA , AC : 100mA
Current Resolution 0.1 μA DC
Current Accuracy ±(1% of reading +0.1% of range) ±(1% of reading +0.1% of range)
Output Frequency 50Hz / 60Hz
Test Time 0.3 ~ 999 sec , continue
Ramp Time 0.1 ~ 99.9 sec, Off
Fall Time 0.1 ~ 99.9 sec, Off
Waveform Sine wave
Insulation Resistance Test
Output Voltage DC : 0.05 ~ 1kV
Voltage Resolution 2V
Voltage Accuracy ±(2% of reading +0.5% of range) ±(2% of reading +0.5% of range)
IR Range 1MΩ ~ 50GΩ
Resistance Resolution 0.1MΩ
Resistance Accuracy 5% typical
Ground Bond Test
Output Current AC : 1 ~ 30A (40A, 60A optional) AC : 3 ~ 40A (60A optional)
Current Accuracy ±(1% of reading +0.2% of range) ±(1% of reading +0.2% of range)
GR Range 10mΩ ~ 510mΩ
Resistance Resolution 0.1mΩ
Resistance Accuracy ± (1% of reading + 0.1% of full scale) ± (1% of reading + 0.1% of full scale)
Test Method 4 wires
Flashover Detection
Setting Mode Programmable setting
Detection Current AC : 20mA, DC : 10mA
Secure Protection Function
Ground Fault Interrupt --- 0.5mA ±0.25mA AC
Floating Output --- 3.5mA, front output only
Panel Operation Lock Present password
Interlock YES
GO/NG Judgment Window
Indication,Alarm GO: Short sound,Green LED; NG: Long sound, Red LED
Data Hold Least tests data memories
Memory Storage 50 setups with up to 100 groups recall
Interface
9pin D-sub I/O control / RS232 / GPIB (Optional)
General
Operation Environment Temperature : 0°C ~ 40°C, Humidity : ± 80 % RH
Power Consumption No load : < 100 W With rated load : 800 W No load : < 100W Rated load : 1000W Maximum load : 1200W
Power Requirements 90~132Vac or 198~264Vac, 47~66Hz
Weight Approx. 20KG(19032) 25KG (19032-P)

 

Hipot Electrical Safety Ground Bond Analyzer

19572 AC current output 45A
Wide resistance range: 10 ~ 510mΩ
UL, CSA, TUV, VDE, CE

 

Model 19572
Mode Ground Bond
Grounding Resistance Test
Output Current AC : 3~45 A
Load Regulation 1% + 0.3 A
Resolution 3~30A 0.01A / 30.1~45A 0.1A
Current Accuracy ±(1.5% of setting + 0.5% of full scale)
Output Frequency 50Hz / 60Hz
Resistance Range 0.1 ~ 510 mΩ
Resistance Resolution R display counts/ I display counts) >0.2, Resolution : 1mΩ
(R display counts/ I display counts) <0.2, Resolution : 0.1mΩ
Resistance Accuracy ±(2% of reading + 0.5% of full scale)
Offset - A predetermined value can be substracted from the measured value and the result of substraction can be displayed.
- The result of subtraction can be compared with a Good/No Good judgement reference value, and the result of the comparison can be used for the judgement.
Offset Range 0 ~ 100 mΩ
Test Time 0.5 ~ 999.0 sec., continue
Waveform Sine Wave
GO/NOGO Judgment - A no-good judgement is made when a resistance greater that the high limit value is detected.
- A no-good judgment is made whne the output current is cutout and a no-good alarm signal is delivered.
- If no abnormal state is detected during the test time, a good judgment is made and a good signal is delivered.
Limit HI-LIMIT : 0.1 ~ 510 mΩ
LOW-LIMIT: OFF, 0.1 ~ HI-LIMIT Value 510 mΩ MAX.
General
Operation Environment Temperature : 0°C~40°C (32° to 104° F), <80% R.H
Power Consumption No load: < 100VA; with rated load: 880W MAX.
Power Requirements 100V / 120V / 220V-10% / 240V-10% ~ +5%
Weight < 16 kg

Signal Test - new and used test equipment
Signal Test Inc

For more information about other electronic test equipment, please visit:
   Hipot Electrical Safety Ground Bond Analyzer
Electronic Workstation Storage Wall System Esd Static Protection Antistatic Control Products
Rigol Oscilloscope Function Arbitrary Wave Form Generator Universal Ic Gang Device Eeprom Memory Flash Programmer

 

 
Presented by: San Diego Online DESIGNED BY  san diego restaurants and San Diego Online.© Copyrights 2009,  by San Diego Online.
If you are interested in listing your business, web site design, or you have any comments, Please send an email to info@sdro.com